Effect of cementless acetabular cup geometry on strain distribution and press-fit stability.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 9139104.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Use of cementless acetabular cups, which are slightly larger than the reamed acetabulum, can provide press-fit stability without screws; however, the ideal cup geometry to maximize stability is not clear. Acetabular strain distribution, deformation, and implant stability were studied using an axisymmetric finite-element model, and mechanical stability was assessed by testing lever-out and extraction forces required to displace different cup geometries from foam bones. The implants tested included four nonhemispheric cup geometries and 1- and 2-mm oversized hemispheric geometries. A nonhemispheric cup that provides a gradual transition from a hemisphere at the dome to a larger peripheral dimension appears to maximize peripheral strains and implant stability without increasing overall acetabular deformation as much as a larger oversized hemispheric cup.