Cardiac pacemaker dysfunction in children after thoracic drainage catheter manipulation.

Lobdell, K W; Walters, H L; Hudson, C; Hakimi, M · Ann Thorac Surg · 1997

case_report · Level V

Where this comes from

Abstract

Two children underwent placement of permanent, epicardial-lead, dual-chamber, unipolar pacemaker systems for complete heart block. Postoperatively, both patients demonstrated subcutaneous emphysema-in the area of their pulse generators-temporally related to thoracic catheter manipulation. Acutely, each situation was managed with manual compression of the pulse generator, ascertaining appropriate pacemaker sensing and pacing. Maintenance of compression with pressure dressings, vigilant observation/monitoring, and education of the care givers resulted in satisfactory pacemaker function without invasive intervention.