Investigation of a Lorentz force biomagnetometer.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 9151478.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
This work evaluates an approach to the noninvasive measurement of small ionic current flows by a technique of Lorentz force magnetometry. An instrument was constructed that is basically a very sensitive force-balance that can measure Lorentz forces experienced by ionic currents flowing in small objects when exposed to strong oscillating magnetic fields. For objects that can fit on a microscope slide, the system is sensitive to ion current dipole moments as low as 180 pA-m. Images were made of ionic currents flowing in thin profiles by a process of scanning a localized magnetic field over the object, measuring generated Lorentz forces, and using a computer to reconstruct images. It can be shown that this method of Lorentz magnetometry has an immunity to ambient magnetic noise and has system characteristics that might suggest its possible use in biomagnetometry of small thin specimens.
Medical subject headings
- Magnetics