Surface-dimpled commercially pure titanium implant and bone ingrowth.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 9152001.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The purpose of this investigation was to examine the effect of the surface macrostructure of a dimpled commercially pure titanium (cp Ti) implant on bone ingrowth in vivo by means of histological examination and a push-out test. Cylindrical implants were inserted in one femur of each experimental rabbit and the animals were killed at 1.5, 3 and 13 months after implantation. The femur with the implant of each animal was then examined in a push-out test. The fracture surfaces of the bone-implant interface after the push-out test were examined under light and electron microscopy. It seems that the dimpled cp Ti surface results in the increased retention of the cp Ti implant in bone due to interlocking between vital bone and the dimples.
Medical subject headings
- Bone and Bones
- Osseointegration
- Prostheses and Implants
- Titanium