RBE variation as a function of depth in the 200-MeV proton beam produced at the National Accelerator Centre in Faure (South Africa).

Gueulette, J; Böhm, L; De Coster, B M; Vynckier, S; Octave-Prignot, M; Schreuder, A N; Symons, J E; Jones, D T et al. · Radiother Oncol · 1997

basic_science · Level V

Where this comes from

Abstract

Thorough knowledge of the RBE of clinical proton beams is indispensable for exploiting their full ballistic advantage. Therefore, the RBE of the 200-MeV clinical proton beam produced at the National Accelerator Centre of Faure (South Africa) was measured at different critical points of the depth-dose distribution. RBEs were determined at the initial plateau of the unmodulated and modulated beam (depth in Perspex = 43.5 mm), and at the beginning, middle and end of a 7-cm spread-out Bragg peak (SOBP) (depths in Perspex = 144.5, 165.5 and 191.5 mm, respectively). The biological system was the regeneration of intestinal crypts in mice after irradiation with a single fraction. Using 60Co gamma-rays as the reference, the RBE values (for a gamma-dose of 14.38 Gy corresponding to 10 regenerated crypts) were found equal to 1.16 +/- 0.04, 1.10 +/- 0.03, 1.18 +/- 0.04, 1.12 +/- 0.03 and 1.23 +/- 0.03, respectively. At all depths, RBEs were found to increase slightly (about 4%) with decreasing dose, in the investigated dose range (12-17 Gy). No significant RBE variation with depth was observed, although RBEs in the SOBP were found to average a higher value (1.18 +/- 0.06) than in the entrance plateau (1.13 +/- 0.04). An RBE value slightly larger than the current value of 1.10 should be adopted for clinical application with a 200-MeV proton beam.