Digital edge-detection measurement of polyethylene wear after total hip arthroplasty.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 9160941.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
A novel digital edge-detection computer technique was developed to measure polyethylene wear after total hip arthroplasty. The new method objectively infers the margins of the component by evaluating gradients of gray-scale intensity on digitized images of the radiographs. Compared with previous methods for measurement of wear, digital edge detection substantially reduces observer subjectivity. The technique was validated directly by measuring wear artificially produced by spherical-front milling of polyethylene liners in a benchtop series. Under such conditions, digital edge detection proved 6.4 times more accurate and 7.1 times more reproducible than manual measurement with conventional circular templates. In addition, clinical application of the new digital imaging technique was illustrated in a series of forty-three patients in whom a metal-backed acetabular cup had been inserted with cement. A random-coefficients prediction algorithm was invoked to estimate long-term wear (mean late rate of wear for the cohort, 0.087 millimeter per year at a mean of 118 months after the operation) on the basis of measurements of short-term wear (mean early rate of wear for the cohort, 0.154 millimeter per year at a mean of twenty-four months).
Medical subject headings
- Hip Prosthesis
- Image Processing, Computer-Assisted
- Polyethylenes
- Prosthesis Failure
Anatomy
- hip