An evaluation of the SCOTI device.
prospective_cohort · Level II
Where this comes from
- Record sourced from PubMed, PMID 9165965.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The Sonomatic Confirmation of Tracheal Intubation (SCOTI, Penlon, UK) device was evaluated in the laboratory with cut and uncut tracheal tubes and in 50 adult patients who underwent tracheal and oesophageal intubation with uncut tracheal tubes. We could not configure the SCOTI reliably with cut tracheal tubes. In the clinical study using uncut tubes the SCOTI identified all oesophageal intubations, but was only able to identify 84% (95% CI: 73.8-94.2%) of tracheal intubations. We recommend caution in using this device and further independent evaluation to determine its reliability.