Memory artifact related to selenium-based digital radiography systems.

Chotas, H G; Floyd, C E; Ravin, C E · Radiology · 1997

basic_science · Level V

Where this comes from

Abstract

Digital images acquired on radiography systems with amorphous selenium detectors are susceptible to "memory artifacts" from prior x-ray exposures. In routine clinical use and in a laboratory experiment, artifacts appeared in chest radiographs until the selenium recovered from initial exposure. Memory artifacts were eliminated when 3 minutes or more elapsed between acquisition of a lateral chest radiograph and acquisition of the next radiograph.