Frequency-domain techniques enhance optical mammography: initial clinical results.

Franceschini, M A; Moesta, K T; Fantini, S; Gaida, G; Gratton, E; Jess, H; Mantulin, W W; Seeber, M et al. · Proc Natl Acad Sci U S A · 1997

case_series · Level IV

Where this comes from

Abstract

We present a novel approach to optical mammography and initial clinical results. We have designed and developed a frequency-domain (110-MHz) optical scanner that performs a transillumination raster scan of the female breast in approximately 3 min. The probing light is a dual-wavelength (690 and 810 nm, 10-mW average power), 2-mm-diameter laser beam, and the detection optical fiber is 5 mm in diameter. The ac amplitude and phase data are processed with use of an algorithm that performs edge effect corrections, thereby enhancing image contrast. This contrast enhancement results in a greater tumor detectability compared with simple light intensity images. The optical mammograms are displayed on a computer screen in real time. We present x-ray and optical mammograms from two patients with breast tumors. Our initial clinical results show that the frequency-domain scanner, even at the present stage of development, has the potential to be a useful tool in mammography.