Maintenance of proximal bone mass with an uncemented femoral stem analysis with dual-energy x-ray absorptiometry.
case_series · Level IV
Where this comes from
- Record sourced from PubMed, PMID 9195311.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Bone ingrowth into uncemented femoral implants with proximal porous coatings has been designed to avoid proximal stress shielding and preserve femoral strength. Dual-energy x-ray absorptiometry allows repeated quantitative analysis of anteroposterior scans of the proximal femur. By use of dual-energy x-ray absorptiometry and qualitative radiographic changes, 31 total hip arthroplasties with an individually designed, proximally porous-coated prosthesis were evaluated after surgery and at intervals up to 2 years. All implants appeared to achieve successful bone ingrowth and subsequent remodeling. At the most proximal level around the neck osteotomy, the postoperative loss of bone density at 6 months was -14.5%, which persisted at 24 months with -11.6%. At the level of the distal portion of the porous coating in the lower metaphysis, the density change was -8.7%, but bone had remodeled at 24 months with a change in density of only -1.0% compared with the immediate postoperative scan. With a design that results in reliable proximal ingrowth, this study predicts that after an initial decline in bone density, a positive bone remodeling response occurs that could lead to long-term stable fixation of the femoral implant.