Adhesive composite resins for artificial teeth: a laboratory investigation of bond strength to a cobalt-chromium alloy.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 9199763.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Adhesive resin systems are reported to improve the bond strength between resins and cast cobalt-chromium alloy. This investigation compares the behaviour of three resin systems. Cylinders and beams of cobalt-chromium, with 0.6-mm-diameter retention beads regularly cast onto the bonding surfaces, were air-abraded and ultrasonically cleaned. Resin veneers 4 mm deep on the cylinders and 2 mm on the bars were polymerized by heat and pressure or by light. Specimens were water-stored for 7 or 90 days, including thermocycling between 4, 37 and 60 degrees C, before testing in a Universal Testing Machine to examine the shear bond strength or the effect of the bonded resin spine on the flexural strength of the beams using a three-point bend test. Specimens were examined with an optical microscope to attempt to determine the nature of the failures that occurred. The investigation showed that, overall, the heat- and pressure-cured urethane dimethacrylate resin with and adhesive based upon methacrylic acid performed significantly less well than a conventional acrylic resin, or a hybrid composite resin with and adhesive monomer.
Medical subject headings
- Chromium
- Cobalt
- Dental Alloys
- Dental Cements
- Tooth, Artificial