Silicone-coated polypropylene hollow-fiber oxygenator: experimental evaluation and preliminary clinical use.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 9205175.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
A membrane oxygenator consisting of a microporous polypropylene hollow fiber with a 0.2-microm ultrathin silicone layer (cyclosiloxane) was developed. Animal experimental and preliminary clinical studies evaluated its reliability in bypass procedures. Five 24-hour venoarterial bypass periods were conducted on dogs using the oxygenator (group A). In 5 controls, bypass periods were conducted using the same oxygenator without silicone coating (group B). As a preliminary clinical study, 14 patients underwent cardiopulmonary bypass with the silicone-coated oxygenator. Eight to 16 hours (mean, 12.2 hours) after initiation of bypass, plasma leakage occurred in all group B animals, but none in group A. The O2 and CO2 transfer rates after 24 hours in group A were significantly higher than at termination of bypass in group B (p < 0.005 and p < 0.03, respectively). Scanning electron microscopy of silicone-coated fibers after 24 hours of bypass revealed no damage to the silicone coating of the polypropylene hollow fibers. In the clinical study, the oxygenator showed good gas transfer, acceptable pressure loss, low hemolysis, and good durability. This oxygenator is more durable and offers greater gas transfer capabilities than the previous generation of oxygenators.
Medical subject headings
- Extracorporeal Membrane Oxygenation
- Materials Testing
- Oxygenators
- Polypropylenes