A silicon bidirectional flow sensor for measuring respiratory flow.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 9214800.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
We describe a solid-state, silicon integrated, bidirectional flow sensor for respiratory applications. The sensor is a thermal vector sensor. The electronic circuitry for obtaining bidirectional sensitivity is presented together with actual application to a healthy volunteer put on mechanical ventilation. The sensor's input flow range is from -60 to +60 L/min, and its rise-time is < or = 40 ms and fall-time is < or = 60 ms. The effect of changes in gas composition as used in mechanically ventilated patients on the sensor output signal are estimated to be less than 2%. The temperature sensitivity is about -1.5% per degree Celsius.
Medical subject headings
- Pulmonary Ventilation
- Silicon
- Ventilators, Mechanical