A silicon bidirectional flow sensor for measuring respiratory flow.

van Putten, M J; van Putten, M H; van Putten, A F; Pompe, J C; Bruining, H A · IEEE Trans Biomed Eng · 1997

basic_science · Level V

Where this comes from

Abstract

We describe a solid-state, silicon integrated, bidirectional flow sensor for respiratory applications. The sensor is a thermal vector sensor. The electronic circuitry for obtaining bidirectional sensitivity is presented together with actual application to a healthy volunteer put on mechanical ventilation. The sensor's input flow range is from -60 to +60 L/min, and its rise-time is < or = 40 ms and fall-time is < or = 60 ms. The effect of changes in gas composition as used in mechanically ventilated patients on the sensor output signal are estimated to be less than 2%. The temperature sensitivity is about -1.5% per degree Celsius.

Medical subject headings