Reproducibility of MEG auditory evoked field source localizations in normal human subjects using a seven-channel gradiometer.

Teale, P; Goldstein, L; Reite, M; Sheeder, J; Richardson, D; Edrich, J; Zimmerman, J E · IEEE Trans Biomed Eng · 1996

basic_science · Level V

Where this comes from

Abstract

Magnetoencephalographic (MEG) auditory evoked fields (EF) were recorded from 12 normal adult subjects over both hemispheres on two separate occasions at least one week apart using a seven-channel second-order gradiometer. Stimuli were computer-generated at 25-msec duration, 1 kHz tone pips. Responses to 100 stimuli were averaged, and source estimates with confidence intervals were computed, for the 100-msec latency auditory EF component, termed M100. Root-mean-squared (rms) differences in x, y, and z locations were approximately 0.7 cm on the two occasions; strength and orientation differences were 18 nA-m and 11 degrees, respectively. This spatial accuracy using a seven-channel instrument, compares favorably with other currently available technologies for localization of brain function.