The time-sequenced adaptive filter for analysis of cardiac arrhythmias in intraventricular electrograms.
other · Level V
Where this comes from
- Record sourced from PubMed, PMID 9216153.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Implantable antitachycardia devices rely upon schemes for detecting cardiac arrhythmias which utilize rate and its variations; yet rate parameters often identify nonpathologic tachycardias as potentially dangerous and deliver unwarranted therapy. I have developed a predictive filter based upon the time-sequenced adaptive algorithm to be used as a supplement to rate criteria for detecting and identifying serious arrhythmias. The method does not require a fixed template and is independent of a priori patient information. The algorithm also provides arrhythmia diagnosis immediately at the change in rhythm. Algorithmic parameters were determined based upon a training set of patient data, and performance of the technique was evaluated with a completely new test set of 20 arrhythmia passages. The new algorithm yielded a sensitivity and specificity for ventricular tachycardia of 91% and 82% and for ventricular fibrillation of 71% and 93%. Correlation waveform analysis was used to diagnose the same test set of arrhythmias. It yielded a sensitivity and specificity for ventricular tachycardia of 100% and 67% and for ventricular fibrillation of 50% and 100%.
Medical subject headings
- Arrhythmias, Cardiac
- Electrocardiography
- Signal Processing, Computer-Assisted