Stereotactic localization with magnetic resonance imaging: a phantom study to compare the accuracy obtained using two-dimensional and three-dimensional data acquisitions.

Walton, L; Hampshire, A; Forster, D M; Kemeny, A A · Neurosurgery · 1997

other

Where this comes from

Abstract

To compare the accuracy of stereotactic localization using magnetic resonance imaging (Siemens 1.5-T Magnetom; Siemens, Erlangen, Germany) with two-dimensional and three-dimensional data acquisition techniques. A phantom study was performed in which the coordinates of an array of rods were determined from images in both two-dimensional and three-dimensional studies and compared with measured values in a series of transverse, coronal, and sagittal images. The results demonstrated a distinct advantage in using three-dimensional acquisition; an error greater than 2 mm was identified in only 0.8% of the imaged volume, compared with 12% of the imaged volume in the two-dimensional study. The results indicated that more accurate stereotactic localization is achieved with a three-dimensional acquisition.