Patient dose due to colon examination: dose assessment and results from a survey in The Netherlands.
retrospective_cohort · Level III
Where this comes from
- Record sourced from PubMed, PMID 9240552.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
To determine the effective dose to the patient during radiographic colon examination. The integral dose-area product was measured during colon examination in 1,733 patients aged 18-94 years. The effective dose was estimated from the dose-area product through computer simulations of radiation transport in anthropomorphic phantoms. The relation between patient dose and imaging or radiographic technique was considered. Patient dose from a biphasic colon examination was compared to that from a double-contrast examination. The factors for converting dose-area product to effective dose were 0.29 mSv x Gy(-1) x cm(-2) and 0.27 mSv x Gy(-1) x cm(-2) for the biphasic and the double-contrast studies, respectively. The average dose-area product for the biphasic colon examination was 21 Gy x cm2, of which 13 Gy x cm2 was attributed to the double-contrast views. The average dose-area product was 29 Gy x cm2 (range, 18-53 Gy x cm2); the average effective dose was 4.7 mSv (range, 2.7-8.4 mSv). Careful selection of the radiologic technique resulted in a surprisingly low dose during the biphasic colon examination. It is recommended that additional filtration of at least 0.1-mm copper be applied and that a screen-film combination with a speed class of at least 400 be used. Dose reduction when using digital techniques is often not realized in clinical practice.
Medical subject headings
- Colon
- Radiation Protection