Phantom assessment of lung dose from proton arc therapy.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 9240659.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
To compare resultant lung dose from proton arc therapy of the chest wall to that from electron arc therapy. A 200 MeV proton beam from the Indiana University Cyclotron was range shifted and modulated to provide a spread out Bragg peak extending from the surface to a depth of 4 cm in water. The chest wall of an Alderson Rando phantom was irradiated by this beam, collimated to a 20 x 4 cm field size, while it rotated on a platform at approximately 1 rpm. For comparison, electron arc therapy of the Rando phantom chest wall was similarly performed with 12 MeV electrons and the resultant lung dose measured in each case. Dose-volume histograms for the Rando phantom left lung indicate a reduced volume of irradiated lung for protons at all dose levels and an integral lung dose that is half that for electron arc therapy in the case studied. In addition, a more uniform dose coverage of the target volume was achieved with the proton therapy. This study demonstrates a potential role for proton arc therapy as an alternative to electron arc therapy when lung dose must be minimized.
Medical subject headings
- Lung
- Phantoms, Imaging
- Protons
- Radiometry