Spinal cord compression caused by the rod of a Harrington instrumentation device: a late complication in scoliosis surgery.
case_report · Level V
Where this comes from
- Record sourced from PubMed, PMID 9258642.
- Also identified by PMC identifier 3454621.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
We report a unique complication caused by the rod of a Harrington instrumentation device, which resulted in spinal stenosis and myelopathy. A literature review revealed no previous causes of direct spinal cord impingement caused by the rod of a Harrington device. In this case, years after the initial operation, the rod penetrated the lamina at the junction between a thoracic and a lumbar curve, causing spinal stenosis and myelopathy. We conclude that regular control of the position of the device and awareness of possible late neurological complications are necessary to identify such complications as early as possible.
Medical subject headings
- Orthopedic Fixation Devices
- Postoperative Complications
- Scoliosis
- Spinal Cord Compression
- Spinal Fusion
Anatomy
- lumbar spine
- thoracic spine