Variation in defibrillation modes after synchronized cardioversion: a potential hazard.
case_report · Level V
Where this comes from
- Record sourced from PubMed, PMID 9259059.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
We conducted a survey of all defibrillators currently on the European market that offer synchronized cardioversion. The mode after cardioversion can be classed into one of four categories: (1) remain in synchronized mode; (2) defer to asynchronized mode; (3) 'intelligent'; and (4) configurable. The different modes and the lack of standardization present a potential hazard because they cause confusion among the user of the defibrillator. Uniformity should be obtained after broad agreement between clinicians and manufacturers of defibrillators.
Medical subject headings
- Electric Countershock