Long-term follow-up of small (size 20 and 21) Medtronic-Hall aortic valve prostheses.
case_series · Level IV
Where this comes from
- Record sourced from PubMed, PMID 9262587.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Our goal was to study the long-term follow-up of patients having aortic valve replacement and to focus particularly on the patients receiving small prostheses. Four hundred twenty-eight Medtronic-Hall valves were implanted (156 size 20 or 21 mm, 149 size 22 or 23 mm, and 123 size 25 or 27 mm). Group 20-21 had a higher number of female patients, more associated coronary lesions, and more patients with aortic stenosis. The actuarial survival rate at 8 years was 80% for group 20-21, 80% for group 22-23, and 76% for group 25-27 (p = not significant). In group 20-21, the actuarial event-free rates at 8 years were as follows: thromboembolic complications, 94%; prosthetic valve endocarditis, 99%; reoperation, 98%; and hemorrhagic complications, 78%. The only factors of prognostic value in this group were age and associated coronary lesions. The durable nature of the results obtained with the Medtronic-Hall 20- and 21-mm prostheses compared with large-diameter prostheses allows the use of a simple and reliable surgical technique and should mean that indications for ring enlargement become rare.
Medical subject headings
- Aortic Valve
- Heart Valve Prosthesis