Long-term mechanical reliability of AMS 700 series inflatable penile prostheses: comparison of CX/CXM and Ultrex cylinders.
retrospective_cohort · Level III
Where this comes from
- Record sourced from PubMed, PMID 9302130.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Recently, we have noted an increasing incidence of revisions being performed in patients implanted with the length and girth expanding AMS 700 Ultrex* inflatable penile prosthesis. This observation prompted us to compare the long-term mechanical reliability of the AMS Ultrex inflatable penile prosthesis versus the girth-expanding AMS 700 CX* or CXM* inflatable penile prosthesis in men with organic erectile dysfunction. Using chart review, mailed questionnaires and telephone interviews, we obtained accurate followup on 111 of 142 (78.2%) patients with CX/CXM implanted between June 1986 and September 1995, and on 152 of 179 (84.9%) patients implanted with Ultrex between October 1989 and September 1995. The CX/CXM and Ultrex groups were compared with regard to 3 end points: 1) mechanical failure caused by any malfunctioning component, 2) device failure caused by any cylinder complication and 3) cylinder aneurysms/leaks. Followup ranged from 1.0 to 112.0 months for the CX/CXM group (mean 47.2 months), and 0.7 to 71.5 months for the Ultrex group (mean 34.4 months). CX/CXM versus Ultrex group comparison demonstrated 10 CX/CXM mechanical failures (9.0%) versus 26 Ultrex failures (17.1%), p = 0.001; 5 CX/CXM cylinder complications (4.5%) versus 13 Ultrex cylinder complications (8.6%), p = 0.0292; and 3 CX/CXM cylinder aneurysms/leaks (2.7%) versus 9 in the Ultrex group (5.9%), p = 0.0162. Kaplan-Meier estimates demonstrated significantly decreased mechanical survival in all 3 categories for Ultrex inflatable penile prosthesis versus CX/CXM inflatable penile prosthesis. Although Ultrex cylinders provide length and girth expansion, Ultrex cylinders exhibit an increased mechanical failure rate at shorter followup compared with CX/CXM cylinders. This increased propensity for Ultrex cylinder problems should be closely monitored.
Medical subject headings
- Penile Prosthesis
- Prosthesis Failure