Biomechanical evaluation of polylactide absorbable screws used for syndesmosis injury repair.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 9347298.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Simulated syndesmosis injuries were created in 12 fresh-frozen, below-knee cadaver specimens. Six specimens were repaired with a 4.5 mm stainless steel screw, and six were repaired with a 4.5 mm polylactide screw. Three specimens of each group were tested in load to failure by axially loading with 1400 N and externally rotating to 90 degrees. Three specimens in each group underwent fatigue testing by axially loading with 700 N and applying 2.5 N-m of torque for 57,700 cycles. Radiographs and computed tomography scans were evaluated. None of the screws broke or failed. Similar load to failure was noted in polylactide and control groups. Fatigue testing revealed no significant change in stiffness. No significant screw damage was evident on radiographic or computed tomography evaluation. The data suggest that a polylactide screw has sufficient fatigue and failure strength to allow for healing of this injury in a clinical situation.
Medical subject headings
- Bone Screws
- Foot Injuries
Anatomy
- foot