Single-incision and single-element array electrode to lower the defibrillation threshold.
case_report · Level V
Where this comes from
- Record sourced from PubMed, PMID 9354555.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Occasional patients have excessive defibrillation energy requirements despite appropriate transvenous defibrillation lead position and the use of biphasic shocks. A single-element subcutaneous array electrode was implanted in 2 patients with a high defibrillation threshold. The array electrode was implanted through the same infraclavicular incision that was used for implantation of the transvenous lead. The defibrillation threshold decreased from 30 J to 15 J and from 24 J to 9 J with the subcutaneous array electrode.
Medical subject headings
- Defibrillators, Implantable