Mechanical obstruction in the anaesthesia delivery-system mimicking severe bronchospasm.
case_report · Level V
Where this comes from
- Record sourced from PubMed, PMID 9370843.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
We present a case where mechanical obstruction in the anaesthesia delivery system caused by plastic wrapping from a filter was misinterpreted as severe bronchospasm. The patient suffered severe hypoxia before the problem was solved by using a free-standing self-expanding ventilation bag. This near-fatal incident emphasises the importance of thorough equipment checking routines, rapid troubleshooting and how equipment failure may be misinterpreted as a medical complication. It also shows how transparent container material can become a medical hazard.
Medical subject headings
- Anesthesia, General
- Anesthesia, Obstetrical
- Bronchial Spasm
- Hypoxia
- Intraoperative Complications