Tourniquet pressures on pediatric patients: a clinical study.
rct · Level II
Where this comes from
- Record sourced from PubMed, PMID 9415909.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
A clinical study was undertaken to evaluate pneumatic tourniquet pressures required for hemostasis in extremity surgery of pediatric patients. Occlusion pressures were measured by Doppler, and tourniquet pressures were set 50 mm above this value. Of 29 cases, 86% were determined to provide adequate hemostasis throughout the procedure. Maximum mean pressures used in the upper and lower extremity groups were 173.4 +/- 11.6 mm Hg (range: 155 to 190 mm Hg) and 176.7 +/- 28.7 mm Hg (range: 140 to 250 mm Hg), respectively, accounting for adjustments made to inadequate initial settings. This study suggests that lower tourniquet pressures than previously used may be needed to maintain adequate hemostasis in pediatric patients.
Medical subject headings
- Blood Loss, Surgical
- Hemostatic Techniques
- Tourniquets