Influence of stem geometry on mechanics of cemented femoral hip components with a proximal bond.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 9420599.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Nonlinear, three-dimensional, finite element models of cemented femoral hip components with a proximal stem-cement bond were developed with use of a Charnley stem geometry and a modified Charnley stem geometry that had a cylindrical cross section over the distal two-thirds of the stem (Distal-Round). Peak tensile stresses in the proximal cement mantle increased 63 and 74% for the Charnley and Distal-Round stems, respectively, when the proximal stem-cement interface was debonded. The shear stresses over the stem-cement interface with a proximal bond were 29% larger for the Distal-Round stem than for the Charnley stem. After the proximal stem-cement interface was debonded, the peak tensile stresses in the cement mantle were 15% larger for the Distal-Round stem than for the Charnley stem. The results illustrate that stresses within the proximal cement mantle could be substantially reduced for both Charnley and Distal-Round stems through use of a proximal stem-cement bond. However, the risk of debonding may be higher for the Distal-Round stem because of increased shear stresses, and once debonded the risk of further loosening due to failure of the cement mantle would also be higher for the Distal-Round stem.
Medical subject headings
- Bone Cements
- Femur
- Hip Prosthesis
Anatomy
- femur
- hip