Percutaneous balloon-assisted thrombectomy: preliminary in vivo results with an expandable vascular sheath system.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 9457198.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
To test an expandable vascular sheath system for percutaneous transcatheter balloon-assisted thrombectomy in vivo. A 9-F expandable vascular sheath system was used. The system consisted of a sheath with a deployable funnel configuration and a 3-F balloon catheter used coaxially. Thrombus was created in the inferior vena cava of 24 rabbits. Thrombus with a mean weight +/- 1 standard deviation of 1,028 mg +/- 67 was removed percutaneously with the expandable sheath system at 2 days (n = 16) and 7 days (n = 8) after formation; 7-day patency was examined after removal of 2-day-old thrombus (n = 8). Histologic examination was performed in the inferior vena cava. In all cases of 2-day-old occlusion, complete patency of the inferior vena cava was achieved. In cases of 7-day-old occlusion, patency could be completely restored in four (50%) cases; in two (25%) cases each, there was partial removal or no thrombus could be extracted. Histologic findings were comparable with those observed after balloon embolectomy. Percutaneous transcatheter balloon-assisted embolectomy is feasible with the expandable vascular sheath system.
Medical subject headings
- Thrombectomy
- Thrombosis
- Vena Cava, Inferior