Correlation of wrist ligamentotaxis with carpal distraction: implications for external fixation.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 9471076.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Cadaver forearms were tested to measure carpal bone separation and wrist ligament tension in response to MTS-based incremental wrist distraction. Distraction of 2 mm separated the proximal carpal row from the radius and transmitted an average tension of 8 N. The mid-carpal joint also began to widen at this level of distraction. Distraction of 4 mm resulted in an average tension of 20 N. At this level of distraction, radioscaphoid separation started to exceed radiolunate separation. A transition from low- to high-stiffness response was observed over a range of 4-8 mm distraction for the 12 specimens tested, with an average tension of 80 N associated with 8 mm distraction. Average values of carpal height ratio, revised carpal height ratio, and carpal height index were found to be poor indicators of distraction, owing to their high variability between specimens.
Medical subject headings
- Carpal Bones
- Ligaments, Articular
- Wrist Joint
Anatomy
- wrist