Polyethylene wear and periprosthetic osteolysis in metal-backed acetabular components with cylindrical liners.
retrospective_cohort · Level III
Where this comes from
- Record sourced from PubMed, PMID 9493531.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
A retrospective clinical and radiographic analysis was performed in 27 hips (23 patients), in which a threaded, porous-coated acetabular component with cylindrical liner was implanted (Sintered T-Tab-ST cup, Biomet, Warsaw, IN). Sixteen hips had cemented stems; 11 had cementless, circumferentially porous-coated stems. The mean clinical and radiographic follow-up period was 68 months. The mean patient age was 49 years; the mean patient weight was 68 kg. The mean rates of linear and volumetric wear were 0.25 mm/y and 149 mm3/y, respectively. Osteolysis occurred in 51.8% (14 hips). Wear rate and young age correlated with osteolysis (P = .0002 and P = .01, respectively). There were no cases of distal femoral osteolysis.
Medical subject headings
- Acetabulum
- Arthroplasty, Replacement, Hip
- Chromium Alloys
- Osteolysis
- Polyethylenes
Anatomy
- hip
- femur