Catastrophic failure of the acetabular component in a ceramic-polyethylene bearing total hip arthroplasty.
case_report · Level V
Where this comes from
- Record sourced from PubMed, PMID 9493548.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Recent research in total hip arthroplasty has focused on attempts to decrease wear at the femoral head-acetabulum articulation, to limit the production of debris that is believed to lead to osteolysis and prosthetic loosening. The use of ceramic-on-polyethylene bearing surfaces has been reported to produce lower wear rates and therefore may increase the life expectancy of the joint arthroplasty. Problems with this bearing have been reported to be due to ceramic femoral head fracture. Reported here are 2 cases of catastrophic failure of total hip arthroplasties, involving a ceramic femoral head, caused by failure of the polyethylene acetabular liner, with subsequent penetration of the femoral head through the acetabular shell.
Medical subject headings
- Acetabulum
- Arthroplasty, Replacement, Hip
- Ceramics
- Polyethylenes
- Prosthesis Failure