Magnification survey and spot view mammography with a new microfocus X-ray unit: detail resolution and radiation exposure.
other
Where this comes from
- Record sourced from PubMed, PMID 9510570.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The aim of our study was to evaluate a mammography unit capable of magnification of up to fourfold at an equivalent or lower dose than with current systems. A prototype mammography tube with an electron-beam-focusing technology resulting in a focal spot size of 40-120 micro m was combined with a highly intensifying screen-film system. To evaluate contrast-detail resolution, phantom radiographs were performed with the prototype magnification mammography system using a magnification factor of 1.7 for survey views and a magnification factor of 4.0 for spot views. They were compared with unmagnified survey views and magnification spot views (magnification factor 1.9) of a state-of-the-art mammography system. The radiation exposure was measured and mean glandular doses were calculated. The contrast-detail resolution with both prototype (m = 1.7) and conventional (m = 1.1) survey views was equivalent while the entrance dose and the mean glandular dose were approximately 50 % lower with the prototype. For spot views, the contrast-detail resolution was substantially higher for the prototype than for conventional magnification while the dose was equivalent. Dose reduction and improved detail resolution are possible with this new technology.
Medical subject headings
- Mammography
- Radiation Dosage
- Radiographic Image Enhancement
- Radiographic Magnification