INR reporting in Canadian medical laboratories: an update. Thrombosis Interest Group of Canada.

Brigden, M L; Johnston, M J · Am J Clin Pathol · 1998

Where this comes from

Abstract

A written survey of all licensed medical laboratories performing coagulation testing in Canada was undertaken in 1996 as a follow-up to a study performed in 1992. The an overall response rate was 77.4% of the laboratories surveyed. During the 4-year period between surveys, the percentage of responding laboratories using some format for international normalized ratio (INR) reporting increased from 57% to 98%. The mean International Sensitivity Index (ISI) of the various thromboplastins used had decreased from 2.07 to 1.63. However, only 35% of responding laboratories used a thromboplastin with an ISI of 1.2 or less as recommended by the American College of Chest Physicians. In 92% of cases, the ISI was said to be specified for the type of instrument, ie, optical density vs mechanical. However, in only 52% of cases was the ISI known to be specific for the particular reagent-analyzer combination used in the individual laboratory. Nine percent of the laboratories were still using an inappropriate denominator for the INR calculation. While 87% of laboratories quoted a therapeutic INR range of 2.0 to 3.0 for venous thromboembolism, there was much greater variation in the range reported for patients undergoing heart valve replacement. The situation in Canada has improved in using the INR system for prothrombin time reporting. However, additional education is required, as is greater use of thromboplastins with a lower ISI.