Effect of locking mechanism on fluid and particle flow through modular acetabular components.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 9590635.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Six modular acetabular components were evaluated to determine whether screw holes in the metal shell offer a route for fluid and debris into the acetabular bone stock. A 56-mm acetabular shell for each trial was mounted to a sealed chamber and loaded at a 25 degrees angle under axial loads of 270-2,700 N and +/- 2.5-N-m torsional load. Polystyrene microspheres (average diameter, 0.5 microm) were placed in double-deionized water at 300 mmH2O pressure in a sealed chamber above the component. The only channel between the fluid above and the collecting chamber below was through the cup-liner interface and 1 screw hole. Fluid and debris in the collecting chamber were harvested after 1,000,000 cycles. The collected sample was filtered through a 0.2-microm-pore filter and analyzed under electron microscopy for evidence of microspheres. Water and polystyrene microspheres were isolated in the collecting chamber for all trials except the Reflection cup (Smith & Nephew Orthopaedics, Memphis, TN) with a screw hole cover and the Micro-Seal cup (Whiteside Biomechanics, St. Louis, MO) with a peripheral seal. A screw placed in the screw hole of the Reflection cup failed to seal the interface. The peripheral seal around the rim of the Micro-Seal polyethylene prevented fluid and particle flow between the metal shell and polyethylene liner.
Medical subject headings
- Joint Prosthesis
- Osteolysis
Anatomy
- hip
- pelvis