2-Methoxyisobutylisonitrile probe during parathyroid surgery: tool or gadget?
case_control · Level III
Where this comes from
- Record sourced from PubMed, PMID 9597920.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The success of parathyroid surgery is determined by the identification and removal of all parathyroid tumors. Parathyroid tumors accumulate and retain 2-methoxyisobutylisonitrile (MIBI) labeled with technetium-99m. Intravenous injection of this radiopharmacon prior to parathyroid surgery allows identification of parathyroid tumors with a hand-held gamma detector. To assess the value of this technique, a case-control study was performed with 62 patient having nuclear-guided parathyroidectomy and 60 patients having conventional parathyroid explorations. The sensitivity rates of the MIBI probe in single and multiple gland disease were 84.6% and 63.0%, respectively. Rates of success, temporary and permanent hypoparathyroidism, and injury of the recurrent laryngeal nerve were similar in patients who underwent probe-guided surgery and those who had conventional surgery. In conclusion, although the MIBI probe appears to be a valuable tool in parathyroid surgery, its use has not improved the outcome of such surgery at our institution.
Medical subject headings
- Parathyroid Neoplasms
- Parathyroidectomy
- Technetium Tc 99m Sestamibi