2-Methoxyisobutylisonitrile probe during parathyroid surgery: tool or gadget?

Bonjer, H J; Bruining, H A; Pols, H A; de Herder, W W; Proye, C A; Carnaille, B M; Mohammedamin, R S; Steyerberg, E W et al. · World J Surg · 1998

case_control · Level III

Where this comes from

Abstract

The success of parathyroid surgery is determined by the identification and removal of all parathyroid tumors. Parathyroid tumors accumulate and retain 2-methoxyisobutylisonitrile (MIBI) labeled with technetium-99m. Intravenous injection of this radiopharmacon prior to parathyroid surgery allows identification of parathyroid tumors with a hand-held gamma detector. To assess the value of this technique, a case-control study was performed with 62 patient having nuclear-guided parathyroidectomy and 60 patients having conventional parathyroid explorations. The sensitivity rates of the MIBI probe in single and multiple gland disease were 84.6% and 63.0%, respectively. Rates of success, temporary and permanent hypoparathyroidism, and injury of the recurrent laryngeal nerve were similar in patients who underwent probe-guided surgery and those who had conventional surgery. In conclusion, although the MIBI probe appears to be a valuable tool in parathyroid surgery, its use has not improved the outcome of such surgery at our institution.