Evaluation of a modified needle for small joint biopsies.
case_series · Level IV
Where this comes from
- Record sourced from PubMed, PMID 9598883.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
To assess the value of a new short biopsy needle modified from the Parker-Pearson needle in obtaining adequate synovial tissue from small joints. The short needle is 2.5 cm in length with a repositioned hooked biopsy notch closer to the end of the needle to allow better specimen retrieval. It can be used to obtain synovial biopsies from small joints of the hand using the usual arthrocentesis approaches. Ten patients underwent 12 synovial biopsies with a success rate of 92%. No complications from the procedure were observed. Samples obtained were adequate for a variety of studies including synovial histopathological evaluation, electron microscopy, and DNA extraction. The new modified short needle expands the spectrum of joints that can be biopsied for diagnostic and research purposes in an outpatient setting.
Medical subject headings
- Arthritis, Rheumatoid
- Biopsy, Needle
- Finger Joint
- Synovial Membrane