Cervical spine: three-dimensional MR imaging with magnetization transfer prepulsed turbo field echo techniques.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 9609910.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
In 11 volunteers, magnetic resonance (MR) imaging of the cervical spine was performed with a magnetization transfer preparatory pulse (prepulsed), three-dimensional Fourier transform, turbo field echo sequence. The effects of flip angle, number of shots, phase-encoding profile order, and magnetization transfer prepulse offset frequency on cerebrospinal fluid-to-cord contrast were evaluated. The contrast was improved by lowering the flip angle, increasing the number of shots, and implementing a magnetization transfer prepulse and linear phase-encoding profile order. Maximum myelographic effect was achieved with the magnetization prepulse (500-Hz frequency offset), 3 degrees flip angle, six shots, and linear phase-encoding profile order.
Medical subject headings
- Cervical Vertebrae
- Magnetic Resonance Imaging