Occlusion of Fontan fenestrations using the Amplatzer septal occluder.
case_series · Level IV
Where this comes from
- Record sourced from PubMed, PMID 9616344.
- Also identified by PMC identifier 1728652.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
To evaluate the efficacy and safety of the Amplatzer septal occluder device for occlusion of Fontan fenestrations. Five children aged 5-10 years who had undergone a fenestrated Fontan operation. Tertiary paediatric cardiology centre. Each patient had right and left heart catheterisation to assess haemodynamic suitability for fenestration closure. Sizing of the defect was achieved with a balloon wedge catheter and transoesphageal echocardiography. Transcatheter occlusion of the fenestration was accomplished using a 4 mm device in three patients, and 5 mm or 9 mm devices in the other two patients. Residual shunting following occlusion was assessed using angiography and echocardiography. 100% occlusion rate of the fenestration was achieved in all patients. No complications or device failures were seen during the three month follow up period. The Amplatzer septal occluder device is safe, and effectively occludes the Fontan fenestration.
Medical subject headings
- Fontan Procedure