Comparison of Multi-Test device skin testing and modified RAST results.
case_control · Level III
Where this comes from
- Record sourced from PubMed, PMID 9627239.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The initial purpose of this study was to determine the potential correlation between allergy test results obtained with the Multi-Test skin testing method and the radioallergosorbent test (RAST) blood test (used as a "standard"). Twenty patients with a history and physical examination findings suggestive of inhalant allergy underwent both a Multi-Test system screen (14 antigens plus histamine and glycerine controls) and RAST testing. The relationship between wheal size and Multi-Test system grade for each antigen and the corresponding RAST class was studied. The correlation between positive Multi-Test system and RAST results was poor, with an average agreement by antigen of 56.26% and overall agreement of 67.86%. However, the overall agreement between negative Multi-Test system results (< or =1+) and negative RAST results (< or =class 1) was 95.15%, with an average agreement by antigen of 83.99%. On the basis of results of this preliminary study, it appears that a negative Multi-Test system result indicates that significant inhalant allergy is unlikely, whereas a positive Multi-Test system result necessitates follow-up with more definitive testing by additional skin testing or RAST.
Medical subject headings
- Hypersensitivity
- Radioallergosorbent Test
- Skin Tests