Moments and forces during pedicle screw insertion. In vitro and in vivo measurements.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 9636975.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Moments and forces during pedicle screw insertion were measured in vivo and in vitro and were correlated to several parameters of the screw-bone interface. To compare the in vitro and in vivo screw insertion loads and to relate these measurements to bone mineral density, pedicle size, and other screw parameters (material, diameter). The in vitro screw insertion torque has been correlated to the screw pullout forces and the number of cycles to ultimate interface failure. However, there are no comparable in vivo data. One hundred three pedicle screws were included in the study, 43 in vivo and 60 in vitro. Duel-energy x-ray absorptiometry boen mineral density data were available for 20 in vivo and 32 in vitro specimens. A custom-made sterilizable six-axis load cell was integrated into a torque wrench, enabling the recording of the applied moments and forces during screw insertion. Statistical analysis was performed to detect differences and correlations. The mean in vivo insertion torque (1.29 Nm) was significantly greater than the in vitro value (0.67 Nm). The linear correlation between insertion torque and bone mineral density was significant for the in vitro data but not for the in vivo data. No correlation was observed between insertion torque and pedicle diameter. Two patterns of torque were observed during the insertion process. There is a significant difference between the insertion loads measured in vivo and those measured in vitro. Additional research is needed to verify whether this method provides an indication of screw fixation quality.
Medical subject headings
- Bone Screws
- Lumbar Vertebrae
Anatomy
- lumbar spine