A study of printed spiral coils for neuroprosthetic transcranial telemetry applications.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 9644895.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
We have explored the use of printed spiral coils (PSC's) for neuroprosthetic transcranial telemetry applications. We fabricated two-dimensional PSC's on a thin (25 microns) polyimide substrate using copper (35 microns) as a conducting material. All the coils had a fixed inner diameter of 1.0 cm. We fabricated two sets of coils. One set of coils consisted of 2- to 5-turn circular and square spiral coils and had different trace widths (W), different spacings (S) between adjacent traces, and different outer diameters. The other set of coils consisted of 5-turn circular spiral coils and had fixed inner and outer diameters but different W to S ratios. We measured loss resistances (Rs and Rp) and quality factors (Q) of these coils at different resonating frequencies in the range of 5-40 MHz. Over this frequency range, we observed that for fixed inner and outer diameters, the coil with the largest W achieved the lowest Rs and the highest Rp and Q. These electrical properties and the fact that these coils can conform to the complex convoluted cortical surface suggest that a PSC [15] can provide a viable alternative to a conventional wire-wound coil for neuroprosthetic transcranial telemetry applications.
Medical subject headings
- Electrodes, Implanted
- Microelectrodes