The forces in the distal femur and the knee during walking and other activities measured by telemetry.
case_report · Level V
Where this comes from
- Record sourced from PubMed, PMID 9645524.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The forces and moments in the shaft of a distal femoral replacement were measured by telemetry for a subject during different activities, and calculations were then made of the forces at the knee. The axial force showed a small peak at heel-strike followed by two main peaks during stance. In level walking, the peak axial force was between 1,487 and 1,718 N (2.2-2.5 BW), the peak shear force was 269-368 N (0.4-0.54 BW) directed anteriorly on the tibia, the peak axial torque was 7 Nm internal, while the patellofemoral force was 466-571 N. The highest axial force was recorded for descending stairs (2.8 BW). Standing on one leg produced 2.4 BW, while lying supine and raising the leg produced 1.7 BW. The data produced may resemble that of a normal subject, and has application to basic joint mechanics, to joint reconstruction, and to total knee replacement design and evaluation.
Medical subject headings
- Knee Joint
- Knee Prosthesis
- Walking
Anatomy
- femur
- knee