Clinical application of a flat-panel X-ray detector based on amorphous silicon technology: image quality and potential for radiation dose reduction in skeletal radiography.
prospective_cohort · Level II
Where this comes from
- Record sourced from PubMed, PMID 9648757.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The purpose of this study was to compare images obtained with a self-scanning, flat-panel X-ray detector based on amorphous silicon technology with conventional screen-film radiographs and to evaluate the possibility of radiation dose reduction in skeletal radiography. One hundred twenty patients were examined prospectively using a conventional screen-film system (speed, 400; detector dose, 2.5 microGy) and a prototype digital amorphous silicon detector (simulated speed, 400 and 800, n = 120; simulated speed, 1600, n = 40). The resulting 400 images were evaluated independently by six radiologists using a subjective five-point preference scale that rated overexposure, underexposure, contrast resolution, spatial resolution, and soft-tissue presentation. Image quality was ranked on a scale from 0 to 10 according to subjective criteria. Statistical significance of differences was determined using Student's t test and confidence intervals (95% confidence level). Comparison of conventional radiographs with digital images revealed a statistically significant preference for the digital system for soft-tissue presentation (speed 400 and 800) and visualization of osteoarthrotic changes (speed 400). A small but statistically significant preference for conventional images was found with respect to contrast and spatial resolution when digital speed was 800 or 1600; and the visibility of arthrosis at digital speed 1600, osteolysis at digital speed 800 and 1600, and fractures at digital speed 1600. The amorphous silicon-based system with a simulated speed of 400 provided images equivalent to screen-film radiographs. For clinical tasks such as routine follow-up studies, assessment of instability, or orthopedic measurements, a radiation dose reduction of up to 75% may be possible.
Medical subject headings
- Bone and Bones
- Technology, Radiologic