This paper describes a method for matching high-temperature superconducting nuclear magnetic resonance (NMR) signal pickup coils to field-effect transistor (FET) preamplifiers using amplifier-specific noise figure and gain parameters to obtain a matching compromise that improves gain flatness for a given permissible noise figure. A technique for obtaining these parameters is also described in an appendix.
Medical subject headings
- Hot Temperature
- Magnetic Resonance Spectroscopy
- Microscopy
- Signal Processing, Computer-Assisted