Initial stability comparison of modular hip implants in synthetic femurs.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 9731670.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Synthetic femurs were used to assess the initial bone-implant interface stability of three total hip systems: Wright Medical Technology's Infinity smooth trochanteric module (S-TM), Infinity porous-coated trochanteric module (PC-TM), and Johnson and Johnson S-ROM with a porous surface. The hips were implanted into synthetic femurs, rigidly fixed, and subjected to internal rotation and cyclic, axial compressive loads. The results showed that all three implants achieved good initial implant stability and would be expected to permit bone ingrowth. The porous-coated implants showed greater initial implant stability with less axial micromotion compared with the smooth implants. This finding suggests that surface texture plays a role in initial stability of uncemented prostheses if the bone behaves similar to the material used in this study.