Real-time bronchoscope tip localization enables three-dimensional CT image guidance for transbronchial needle aspiration in swine.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 9824022.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
STUDY OBJECTIVE: To determine the feasibility of using real-time bronchoscope position technology coupled with previously acquired three-dimensional CT data to enhance transbronchial needle aspiration (TBNA). DESIGN: Eight swine were given percutaneously created target lesions for TBNA. A miniature position sensor was placed at the tip of a bronchoscope, and real-time position information during bronchoscopy was presented on a monitor simultaneously displaying previously acquired three-dimensional CT data. INTERVENTIONS: TBNA of target lesions and submucosal ink-spot injection of computer-generated targets. MEASUREMENTS AND RESULTS: TBNA specimens revealed successful aspiration of target material. Distances between ink marks made at computer-generated tracheal targets varied, on average (+/- SD), 4.2 mm +/- 2.6 mm from predetermined computer-distance coordinates. CONCLUSION: Real-time bronchoscope position technology coupled with previously acquired CT images may aid with TBNA of nonvisible extrabronchial lesions.
Medical subject headings
- Biopsy, Needle
- Bronchoscopy
- Image Processing, Computer-Assisted
- Radiography, Interventional
- Tomography, X-Ray Computed