Results of cemented metal-backed acetabular components: a 10-year-average follow-up study.
retrospective_cohort · Level III
Where this comes from
- Record sourced from PubMed, PMID 9880177.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The clinical and radiographic results of 86 primary total hip arthroplasties performed in 74 patients from 1983 to 1987 with a cemented metal-backed acetabular component and a cemented collared straight femoral stem with a 32-mm head were reviewed at a mean follow-up of 10.1 years. Seven patients (9.2%) underwent acetabular component revision at a mean of 9.0 years after implantation; an additional 24 components (31.6%) demonstrated evidence of radiographic loosening, resulting in a total failure rate of 40.8%. Periacetabular radiolucencies were noted in Charnley zones at the following rates: 34.2% in zone I, 18.4% in zone II, and 27.6% in zone III. In addition, 18.4% and 38.2% of implants demonstrated evidence of migration and excessive polyethylene wear. Excessively vertical cup placement (>49 degrees inclination) at the time of initial arthroplasty was statistically correlated with polyethylene wear, implant migration, and fixation failure. A trend of increasing implant failure was also noted with decreasing polyethylene liner thickness. Periacetabular cement mantle thickness was not statistically correlated with subsequent component loosening or failure. Results of Kaplan-Meier survivorship analysis using revision as an endpoint showed 93.6% survivorship at 10 years and 88.4% at 12 years. The mean modified Harris hip scores were 46.9 preoperatively and 81.8 at final follow-up. The significant overall rates of radiographic loosening, migration, polyethylene wear, and implant revision confirm the suspected trend of increasing failure rates of cemented metal-backed acetabular components over time.
Medical subject headings
- Arthroplasty, Replacement, Hip
- Bone Cements
- Hip Prosthesis
Anatomy
- hip
- pelvis