Cartilage degeneration in relation to repetitive pressure: case study of a unilateral hip hemiarthroplasty patient.
case_report · Level V
Where this comes from
- Record sourced from PubMed, PMID 9926953.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
In vivo acetabular contact pressures were measured over 32 months in an elderly man with a pressure instrumented hemiarthroplasty. After death, left (hemiarthroplasty) and right (control) acetabula were explanted. Cartilage thickness and degeneration were quantified from magnetic resonance imaging and histological analysis. Highest repetitive in vivo contact pressures during gait (4.5 to 6.5 MPa) were measured in the superior dome of the acetabulum and decreased at a rate of approximately 1 MPa per year after implant (R2 = 0.48, P < .001). Contact pressure magnitudes measured during gait correlated positively with regional histology score (R2 = 0.34, P < .0001) and negatively with cartilage thickness (R2 = 0.35, P < .0001). Although histology scores were typical of early osteoarthritis (histological grade of 4-6), there were no significant differences in overall histology score for the left and right acetabula (P = .23). We conclude that acetabular cartilage degeneration was explained, in part, by repetitive stress, but the degeneration did not appear to be mediated solely by articulation with the metallic endoprosthesis.
Medical subject headings
- Cartilage, Articular
- Hip Prosthesis
Anatomy
- hip