ECG beat detection using filter banks.

Afonso, V X; Tompkins, W J; Nguyen, T Q; Luo, S · IEEE Trans Biomed Eng · 1999

basic_science · Level V

Where this comes from

Abstract

We have designed a multirate digital signal processing algorithm to detect heart beats in the electrocardiogram (ECG). The algorithm incorporates a filter bank (FB) which decomposes the ECG into subbands with uniform frequency bandwidths. The FB-based algorithm enables independent time and frequency analysis to be performed on a signal. Features computed from a set of the subbands and a heuristic detection strategy are used to fuse decisions from multiple one-channel beat detection algorithms. The overall beat detection algorithm has a sensitivity of 99.59% and a positive predictivity of 99.56% against the MIT/BIH database. Furthermore this is a real-time algorithm since its beat detection latency is minimal. The FB-based beat detection algorithm also inherently lends itself to a computationally efficient structure since the detection logic operates at the subband rate. The FB-based structure is potentially useful for performing multiple ECG processing tasks using one set of preprocessing filters.

Medical subject headings